5-38
PER1 BS Level
CONFigure:SEQuence:PER1bslevel <real>
CONFigure:SEQuence:PER1bslevel?
This command requires the installation of the Option C02.
This command is valid when CDMA mode is set to EDVO. This
specifies the amplitude (BS Level) for PER1 measurement in
each band set by the
Band
from -115.0 to -18.0 (dBm) in 0.1 steps. At preset, this is set
to -105.5 (dBm).
PER2 BS Level
CONFigure:SEQuence:PER2bslevel <real>
CONFigure:SEQuence:PER2bslevel?
This command requires the installation of the Option C02.
This command is valid when CDMA mode is set to EDVO. This
specifies the amplitude (BS Level) for PER2 measurement in
each band set by the
Band
from -115 .0 to -18.0 (dBm) in 0.1 steps. At preset, this is
set to -25.0 (dBm).
PER3 BS Level
CONFigure:SEQuence:PER3bslevel <real>
CONFigure:SEQuence:PER3bslevel?
This command requires the installation of the Option C02.
This command is valid when CDMA mode is set to EDVO. This
specifies the amplitude (BS Level) for PER3 measurement in
each band set by the
Band
from -115.0 to -18.0 (dBm) in 0.1 steps. At preset, this is
set to -25.0 (dBm).
RF Signal Output Mode
CONFigure:SEQuence:RFOUtput AUTO|ON
CONFigure:SEQuence:RFOUtput?
N9360A Multi UE Tester CDMA2000 Programming Manual
command. The allowable range is
command. The allowable range is
command. The allowable range is