Programming Command Reference
5-72
Sensitivity/TDSO FER (95%)
CONFigure:CONDition:LIMit:SENSitivity:TDSOFER:<<
95 <real HI>
CONFigure:CONDition:LIMit:SENSitivity:TDSOFER:95
?
This command is valid when CDMA mode is set to MC1X. This
specifies the high limit for Sensitivity/TDSO FER (95%) test in each
band set by the
Band
is from 0.0 to 5.0 (%) in 0.5 steps. At preset, this is set to 0.5
(%).
This query command returns the set high limit of
Sensitivity/TDSO FER (95%) test with real. The range of the value
is same as that of the command parameter.
Sensitivity/TDSO FER (Off)
CONFigure:CONDition:LIMit:SENSitivity:TDSOFER:<<
Off <real HI>
CONFigure:CONDition:LIMit:SENSitivity:TDSOFER:Of
f?
This command is valid when CDMA mode is set to MC1X. This
specifies the high limit for Sensitivity/TDSO FER (Off) test in each
band set by the
Band
is from 0.00 to 100.00 (%) in 0.01 steps. At preset, this is set
to 0.50 (%).
This query command returns the set high limit of
Sensitivity/TDSO FER (Off) test with real. The range of the value is
same as that of the command parameter.
Sensitivity/PER Test 1 (95%)
CONFigure:CONDition:LIMit:SENSitivity:PER1:<<
95 <real HI>
CONFigure:CONDition:LIMit:SENSitivity:PER1:95?
This command requires the installation of the Option C02.
N9360A Multi UE Tester CDMA2000 Programming Manual
command on
page
36. The allowable range
command on
page
36. The allowable range