N9360A Multi UE Tester CDMA2000 Programming Manual
This command sets the amplitude (BS Level) of output signal
from the Tester in each band set by the
page
36, and can be used in Stand-by or Connection status. The
allowable range is from -115.0 to -18.0 (dBm) in 0.1 steps.
At preset, this is set to -75.0 (dBm).
This query command returns the set amplitude of output signal
with real. The range of the value is same as that of the command
parameter.
PER BS Level
TESTs:MANual:PERbslevel <real>
TESTs:MANual:PERbslevel?
This command is valid when CDMA mode is set to EVDO. This
specifies the amplitude (BS Level) for PER measurement in
each band set by the
Band
range is from -115.0 to -18.0 (dBm) in 0.1 steps. At preset,
this is set to -75.0 (dBm).
MAX Power BS Level
TESTs:MANual:MAXpowerbslevel <real>
TESTs:MANual:MAXpowerbslevel?
This command is valid when CDMA mode is set to MC1x or
EVDO. This specifies the amplitude (BS Level) for Max Tx Power
measurement in each band set by the
36. The allowable range is from -115 .0 to -18.0 (dBm) in
page
0.1 steps. At preset, this is set to -104.0 (dBm) for MC1x,
-60.0 (dBm) for EVDO.
Programming Command Reference