Advertisement

5
Programming Command Reference
Sensitivity/PER Test 3 (Off)
CONFigure:CONDition:LIMit:SENSitivity:PER3:<<
Off <real HI>
CONFigure:CONDition:LIMit:SENSitivity:PER3:Off?
This command requires the installation of the Option C02.
This command is valid when CDMA mode is set to EVDO. This
specifies the high limit for Sensitivity/PER Test 3 (Off) test in each
band set by the
Band
command. The allowable range is from
0.00 to 100.00 (%) in 0.01 steps. At preset, this is set to 0.50
(%).
This query command returns the set high limit of Sensitivity/PER
Test 3 (Off) test with real. The range of the value is same as that
of the command parameter.
3GPP system
CONFigure:CONDition:3GPPsystem <int>
CONFigure:CONDition:3GPPsystem?
This command sets 3GPP system. The range is from 1 to 5 in 1
steps. At preset, this is set to 2.
This query command returns the set 3GPP system with integer.
The range of the value is same as that of the command
parameter.
Signaling Pattern
CONFigure:CONDition:SIGNalingpattern <int>
CONFigure:CONDition:SIGNalingpattern?
This command sets a Signaling Pattern. The allowable range is
from 1 to 255 in 1 steps. At preset, this is set to 1.
This query command returns the set Signaling Pattern with
integer. The range of the value is same as that of the command
parameter.
N9360A Multi UE Tester CDMA2000 Programming Manual
5-75