• <real 3> (up to 5 letters) is the measured data ranging
from -99.9 to +99.9 in dBm. If measurement result is out
of range or the Access Probe Power is not measured, the value
+99.9 is returned.
Pilot Strength Measurement
TESTs:MANual:RESults:PLTStrength?
This query command returns the Pilot Strength Measurement
result values, <int 1>, <int 2>, and <real 3>.
• <int 1> (1 digit) is 0 if measurement is completed, or 1 if
measurement is not completed.
• <int 2> (1 digit) returns the same value as <int 1>.
(Judgment of limit value is not executed.)
• <real 3> (up to 5 letters) is the measured data ranging
from -31.5 to 0.0 in dB. If measurement result is out of
range or the Pilot Strength is not measured, the value 99.9
is returned.
RX Power Measurement
TESTs:MANual:RESults:RXPower?
This query command is valid when CDMA mode is set to MC1X.
This query command returns the RX Power Measurement result
values, <int 1>, <int 2>, and <real 3>.
• <int 1> (1 digit) is 0 if measurement result is within the
measurement range, or 1 if measurement result is out of
range or the RX Power is not measured.
• <int 2> (1 digit) is euqal to <int 1>. (Judgment of limit
value is not executed.)
• <real 3> (up to 4 letters) is the measured data ranging
from -120 to -18 in dBm. If any measurement is overflow,
underflow or not made, the value 999 is returned.