This query command returns the setting of FER Confidence with
the value of 95% or OFF.
PER FT Rate
TESTs:MANual:FTRAte 307.2|2457.6|3072.0
TESTs:MANual:FTRAte?
This command requires the installation of the Option C02.
This command is valid when CDMA mode is set to EVDO. This
sets the PER FT Rate. The parameter is 307.2 (Low) or 2457.6
(High). At preset, this is set to 307.2.
This query command returns the setting of PER FT Rate with the
value of 307.2 or 2457.6.
PER Max Packets
TESTs:MANual:PER:MAXPackets <int>
TESTs:MANual:PER:MAXPackets?
This command requires the installation of the Option C02.
This command is valid when CDMA mode is set to EVDO. This
sets the PER Max Packets. The allowable range is from 25 to
10000000 in 1 steps. At preset, this is set to 10000.
This query command returns the setting of PER Max Packets with
integer. The range of the value is same as that of the command
parameter.
PER Confidence
TESTs:MANual:PER:CONFidence 95%|OFF
TESTs:MANual:PER:CONFidence?
This command requires the installation of the Option C02.
This command is valid when CDMA mode is set to EVDO. This
sets the PER Confidence. The parameter is 95% or OFF. At preset,
this is set to 95%.
This query command returns the setting of PER Confidence with
the value of 95% or OFF.
N9360A Multi UE Tester CDMA2000 Programming Manual