N9360A Multi UE Tester CDMA2000 Programming Manual
These query commands return the Time Offset Measurement
result values, <int 1>, <int 2> and <real 3> for each
channels.
• <int 1> (1 digit) is 0 if measurement result is within the
measurement range, or 1 if measurement result is out of
range or Time Offset is not measured.
• <int 2> (1 digit) is the test result for the 0 (pass) or 1 (fail).
• <real 3> (up to 6 letters) is the measured data ranging
from -10.00 to +10.00 in
of range or Time Offset is not measured, the value +99.99 is
returned.
Sensitivity/FER Measurement
TESTs:AUTO:RESults:SENSitivity:FER1?
TESTs:AUTO:RESults:SENSitivity:FER2?
TESTs:AUTO:RESults:SENSitivity:FER3?
TESTs:AUTO:RESults:SENSitivity:FER4?
TESTs:AUTO:RESults:SENSitivity:FER5?
TESTs:AUTO:RESults:SENSitivity:FER6?
These query commands are valid when CDMA mode is set to
MC1X. They return the Sensitivity/FER Measurement result values,
<int 1>, <int 2> and <real 3> for each channels.
• <int 1> (1 digit) is 0 if measurement result is within the
measurement range, or 1 if measurement result is out of
range or Sensitivity/FER is not measured.
• <int 2> (1 digit) is the test result for the 0 (pass) or 1 (fail).
• <real 3> (up to 6 letters) is the measured data ranging from
0.00 to 100.00 %. If measurement result is out of range or
Sensitivity/FER is not measured, the value 999.99 is
returned.
Sensitivity/PER1 Measurement
TESTs:AUTO:RESults:SENSitivity:PER11?
TESTs:AUTO:RESults:SENSitivity:PER12?
TESTs:AUTO:RESults:SENSitivity:PER13?
Programming Command Reference
μ
sec. If measurement result is out
5
5-97