Programming Command Reference
5-126
TX Power Measurement
TESTs:MANual:RESults:TXPower?
This query command returns the TX Power measurement result
values, <int 1>, <int 2>, and <real 3>.
• <int 1> (1 digit) is 0 if measurement result is within the
measurement range, or 1 if measurement result is out of
range or the TX Power is not measured.
• <int 2> (1 digit) is the test result for the 0 (pass) or 1 (fail).
(Judgment of limit value is not executed.)
• <real 3> (up to 6 letters) is the measured data ranging
from -99.99 to +99.99 in dBm. If any measurement is
overflow, underflow or not made, the value +99.99 is
returned.
MAX TX Power Measurement
TESTs:MANual:RESults:TXPower:MAXimum?
This query command returns the MAX TX Power Measurement
result values, <int 1>, <int 2>, and <real 3>.
• <int 1> (1 digit) is 0 if measurement result is within the
measurement range, or 1 if measurement result is out of
range or MAX TX Power is not measured.
• <int 2> (1 digit) is the test result for the pass (0) or fail (1).
• <real 3> (up to 6 letters) is the measured data ranging
from -99.99 to +99.99 in dBm. If measurement result is
out of range or MAX TX Power is not measured, the value
+99.99 is returned.
MIN TX Power Measurement
TESTs:MANual:RESults:TXPower:MINimum?
This query command returns the MIN TX Power Measurement
result values, <int 1>, <int 2>, and <real 3>. Each value
is separated by a comma.
• <int 1> (1 digit) is 0 if measurement result is within the
measurement range, or 1 if measurement result is out of
range or MIN TX Power is not measured.
• <int 2> (1 digit) is the test result for the 0 (pass) or 1 (fail).
N9360A Multi UE Tester CDMA2000 Programming Manual