Installation Note
Configurable Test Set Upgrade Kit
Upgrade Kit Number: N5230-60102
For N5230A/C Standard Test Set, 6 GHz and 13.5 GHz, 2-Port PNA-L
Agilent Part Number: N5230-90004
Printed in USA December 2007
Supersedes: July 2007
© Agilent Technologies, Inc. 2005-2007

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies N5230-60102

  • Page 1 Installation Note Configurable Test Set Upgrade Kit Upgrade Kit Number: N5230-60102 For N5230A/C Standard Test Set, 6 GHz and 13.5 GHz, 2-Port PNA-L Agilent Part Number: N5230-90004 Printed in USA December 2007 Supersedes: July 2007 © Agilent Technologies, Inc. 2005-2007...
  • Page 2: Warranty Statement

    “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987).
  • Page 3: Contacting Agilent

    Contacting Agilent Assistance with test and measurement needs and information on finding a local Agilent office are available on the Web at: http://www.agilent.com/find/assist If you do not have access to the Internet, please contact your Agilent field engineer. In any correspondence or telephone conversation, refer to the Agilent NOTE product by its model number and full serial number.
  • Page 4: Items Included In The Upgrade Kit

    Items Included in the Upgrade Kit Check the contents of your kit against the following list. If any part is missing or damaged, contact Agilent Technologies. Refer to “Contacting Agilent” on page Table 1 Contents of Upgrade Kit N5230-60102 Part Description Desig. Number...
  • Page 5 Installation Procedure for the Upgrade The network analyzer must be in proper working condition prior to installing this option. Any necessary repairs must be made before proceeding with this installation. This installation requires the removal of the analyzer’s protective outer WARNING covers.
  • Page 6 Overview of the Installation Procedure Step 1. Remove the Outer Cover. Step 2. Remove the Front Panel Assembly. Step 3. Remove the Existing Cables. Step 4. Install the Step Attenuators. Step 5. Install the New Cables. Step 6. Replace the Lower Front Panel Overlay. Step 7.
  • Page 7 Step 1. Remove the Outer Cover This procedure is best performed with the analyzer resting on its front handles CAUTION in the vertical position. Do not place the analyzer on its front panel without the handles. This will damage the front panel assemblies. Refer to Figure 2 for this procedure.
  • Page 8: Step 2. Remove The Front Panel Assembly

    Step 2. Remove the Front Panel Assembly Refer to Figure 3 for this procedure. ① 1. With a T-10 TORX driver, remove the eight screws (item ) from the sides of the frame. Before removing the front panel from the analyzer, lift and support the front of CAUTION the analyzer chassis.
  • Page 9 Step 3. Remove the Existing Cables Refer to Figure 4 for this procedure. 1. Place the analyzer bottom-side up on a flat surface. 2. Remove the following cables in the order listed: • W7 A19 SSLAM to A21 test port 1 coupler •...
  • Page 10 Step 4. Install the Step Attenuators Refer to Figure 5 for this procedure. New parts are listed in Table 1 on page ① ② 1. Loosely install two screws (item ) in the side of each attenuator (item ). Insert each screw only a few threads into the attenuator.
  • Page 11 Step 5. Install the New Cables Refer to Figure 6 for the following procedure. New parts are listed in Table 1 on page 1. Install the following cables in the order listed. Use a 5/16-in torque wrench set to 10 in-lbs. •...
  • Page 12 Figure 6 New Cable Installation Installation Note N5230-90004...
  • Page 13 Step 6. Replace the Lower Front Panel Overlay Refer to Figure 7 for this procedure. New parts are listed in Table 1 on page 1. From the back side of the front panel, use a blunt object in one of the cutouts in the lower ①...
  • Page 14 Step 7. Reinstall the Front Panel Assembly and Install the Front Panel Jumpers Before installing the front panel assembly onto the analyzer, lift and support the CAUTION front of the analyzer chassis. Refer to Figure 8 for this procedure. New parts are listed in Table 1 on page 1.
  • Page 15 The previous figure and the figure below show the N5230A front panel and NOTE floppy disk drive. The N5230C front panel has a slightly different appearance and does not include a floppy disk drive. Step 8. Reinstall the Outer Cover This procedure is best performed with the analyzer resting on its front handles CAUTION in the vertical position.
  • Page 16 Step 9. Enable Options 014 and 1E1 Procedure Requirements • The analyzer must be powered up and operating to perform this procedure. • The Network Analyzer program must be running. • A mouse must be connected to the analyzer for this procedure. Option Enable Procedure 1.
  • Page 17 Step 10. Perform Post-Upgrade Adjustments and Calibration Adjustments The following adjustments must be made due to the hardware changes of the analyzer. • source calibration • receiver calibration These adjustments are described in the PNA service guide and in the PNA on-line HELP. A list of equipment required to perform these adjustments can be found at “Equipment Required for Post-Upgrade Adjustments”...
  • Page 18 Installation Note N5230-90004...