N9360A Multi UE Tester CDMA2000 Programming Manual
Frequency Error Measurement
TESTs:AUTO:RESults:FERRor1?
TESTs:AUTO:RESults:FERRor2?
TESTs:AUTO:RESults:FERRor3?
TESTs:AUTO:RESults:FERRor4?
TESTs:AUTO:RESults:FERRor5?
TESTs:AUTO:RESults:FERRor6?
These query commands return the Frequency Error Measurement
result values, <int 1>, <int 2> and <real 3> for each
channels.
• <int 1> (1 digit) is 0 if measurement result is within the
measurement range, or 1 if measurement result is out of
range or Frequency Error is not measured.
• <int 2> (1 digit) is the test result for the 0 (pass) or 1 (fail).
• <real 3> (up to 8 letters) is the measured data ranging
from -10000.0 to +10000.0 in Hz. If measurement result is
out of range or Frequency Error is not measured, the value
+99999.9 is returned.
Rho Measurement
TESTs:AUTO:RESults:RHO1?
TESTs:AUTO:RESults:RHO2?
TESTs:AUTO:RESults:RHO3?
TESTs:AUTO:RESults:RHO4?
TESTs:AUTO:RESults:RHO5?
TESTs:AUTO:RESults:RHO6?
These query commands are valid when CDMA mode is set to
MC1X. They return the Rho Measurement result values, <int 1>,
<int 2> and <real 3> for each channel, when F1R1 or F2R2
is set by the Radio Configuration command.
• <int 1> (1 digit) is "0" if measurement result is within the
measurement range, or "1" if measurement result is out of
range or Rho is not measured.
• <int 2> (1 digit) is the test result for the 0 (pass) or 1 (fail).
Programming Command Reference
5
5-95