Agilent Technologies N9360A Programming Manual page 40

Multi ue tester
Hide thumbs Also See for N9360A:
Table of Contents

Advertisement

4
Programming Command Guidelines
4-12
Flow Control
If the Tester cannot be controlled with the controller caused by
the receiver buffer full of the Tester, use Xon/Xoff flow control.
To use the Xon/Xoff flow control with the Tester, set the Xcontrol
on the [Configuration] screen to
receives the
(decimal value of 19), stop transmitting data.
Xoff
When the controller receives the
resume transmitting data.
Flow control using the control lines (CTS, DSR and DCD) are
not available.
Data Transfer Errors
On the serial interface several type of errors may occur during
data reception. The errors may occur by any of the following
factor.
• Parity error: The parity bit on a receiving character does not
match the parity bit expected by the receiver.
This error is most commonly caused by line
noise. In this case, reducing the cable length or
using a shielded cable may solve this error.
• Framing error: The start bit and stop bit do not match the
timing expectations of the receiver. This
error is most commonly cased the cable line
noise. In this case, reducing the cable length
or using a shielded cable may solve this error.
• Overrun error: A receiver buffer overflows, and received
data is lost. This error is most commonly
caused by too fast communication speed
(Baud Rate). In this case, reduce the
communication speed (Baud Rate).
N9360A Multi UE Tester CDMA2000 Programming Manual
. When the controller
Xon/Xoff
(decimal value of 17),
Xon