Agilent Technologies N9360A Programming Manual page 124

Multi ue tester
Hide thumbs Also See for N9360A:
Table of Contents

Advertisement

5
Programming Command Reference
5-74
Sensitivity/PER Test 2 (Off)
CONFigure:CONDition:LIMit:SENSitivity:PER2:<<
Off <real HI>
CONFigure:CONDition:LIMit:SENSitivity:PER2:Off?
This command requires the installation of the Option C02.
This command is valid when CDMA mode is set to EVDO. This
specifies the high limit for Sensitivity/PER Test 2 (Off) test in each
band set by the
Band
is from 0.00 to 100.00 (%) in 0.01 steps. At preset, this is set
to 0.50 (%).
This query command returns the set high limit of Sensitivity/PER
Test 2 (Off) test with real. The range of the value is same as that
of the command parameter.
Sensitivity/PER Test 3 (95%)
CONFigure:CONDition:LIMit:SENSitivity:PER3:<<
95 <real HI>
CONFigure:CONDition:LIMit:SENSitivity:PER3:95?
This command requires the installation of the Option C02.
This command is valid when CDMA mode is set to EVDO. This
specifies the high limit for Sensitivity/PER Test 3 (95%) test in
each band set by the
Band
range is from 0.0 to 5.0 (%) in 0.5 steps. At preset, this is set
to 0.5 (%).
This query command returns the set high limit of Sensitivity/PER
Test 3 (95%) test with real. The range of the value is same as that
of the command parameter.
N9360A Multi UE Tester CDMA2000 Programming Manual
command on
page
36. The allowable range
command on
page
36. The allowable