Agilent Technologies N9360A Programming Manual page 180

Multi ue tester
Hide thumbs Also See for N9360A:
Table of Contents

Advertisement

5
Programming Command Reference
5-130
Time Offset Measurement
TESTs:MANual:RESults:TOFFset?
This query command returns the Time Offset Measurement result
values, <int 1>, <int 2>, and <real 3>.
• <int 1> (1 digit) is 0 if measurement result is within the
measurement range, or 1 if measurement result is out of
range or Time Offset is not measured.
• <int 2> (1 digit) is the test result for the pass (0) or fail (1).
• <real 3> (up to 6 letters) is the measured data ranging
from -10.00 to +10.00 in
of range or Time Offset is not measured, the value +99.99 is
returned.
Code Domain Power (TX Power) Measurement
TESTs:MANual:RESults:CDPWr:TXPower?
This query command returns the TX Power Measurement result
values, <int 1>, <int 2>, and <real 3> on the [Code
Domain Power] measurement screen. Each value is separated
by a comma.
• <int 1> (1 digit) is 0 if measurement result is within the
measurement range, or 1 if measurement result is out of
range or TX Power is not measured.
• <int 2> (1 digit) is the test result for the pass (0) or fail (1).
(Judgment of limit value is not executed.)
• <real 3> (up to 6 letters) is the measured data ranging
from -99.99 to +99.99 in dBm. If measurement result is
out of range or TX Power is not measured, the value +99.99
is returned.
Code Domain Power Pilot Measurement
TESTs:MANual:RESults:CDPWr:PILot?
This query command returns the Code Domain Power Pilot CH
Measurement result values, <int 1>, <int 2>, and <real 3>.
• <int 1> (1 digit) is 0 if measurement result is within the
measurement range, or 1 if measurement result is out of
range or Code Power Pilot is not measured.
N9360A Multi UE Tester CDMA2000 Programming Manual
μ
sec. If measurement result is out

Advertisement