N9360A Multi UE Tester CDMA2000 Programming Manual
Table 5-21 Power Control Level Parameter and Preset Value (for
Connection)
Parameter
Description
<int>
1 to 99 dB in 1 steps.
The available parameters are limited by the parameter currently
set in the Tester as follows:
If currently set to HOLD, all the parameters (HOLD, UP, DOWN,
CNTUP and CNTDWN) are available.
If currently set to CNTUP, parameters HOLD, DOWN, CNTUP and
CNTDWN are available.
If currently set to CNTDWN, parameters HOLD, UP, CNTUP and
CNTDWN are available.
TESTs:MANual:PLEVel?
This query command returns the set Power Control Level with
integer. The range of the value is same as that of the command
parameter.
FER Max Frames
TESTs:MANual:FER:MAXFrames <int>
TESTs:MANual:FER:MAXFrames?
This command is valid when CDMA mode is set to MC1X. This
sets the number of FER Max Frames. The allowable range is from
25 to 10000000 in 1 steps. At preset, this is set to 10000.
This query command returns the set number of FER Max Frames
with integer. The range of the value is same as that of the
command parameter.
FER Confidence
TESTs:MANual:FER:CONFidence 95%|OFF
TESTs:MANual:FER:CONFidence?
This command is valid when CDMA mode is set to MC1X. This
sets the FER Confidence. The parameter is 95% or OFF. At preset,
this is set to 95%.
Programming Command Reference
Preset value
1
5
5-107