N9360A Multi UE Tester CDMA2000 Programming Manual
• <int 1> (1 digit) is 0 if measurement result is within the
measurement range, or 1 if measurement result is out of
range or TX Power is not measured.
• <int 2> (1 digit) is the test result for the 0 (pass) or 1 (fail).
(Judgment of limit value is not executed.)
• <real 3> (up to 6 letters) is the measured data ranging
from -99.99 to +99.99 in dBm. If measurement result is
out of range or TX Power is not measured, the value 99.99 is
returned.
Code Domain Power Pilot Measurement
TXANalyzer:RESults:CDPWr:PILot?
This query command returns the Code Domain Power Pilot CH
Measurement result values, <int 1>, <int 2>, and <real 3>.
Each value is separated by a comma.
• <int 1> (1 digit) is 0 if measurement result is within the
measurement range, or 1 if measurement result is out of
range or Code Domain Power Pilot CH is not measured.
• <int 2> (1 digit) is the test result for the 0 (pass) or 1 (fail).
(Judgment of limit value is not executed.)
• <real 3> (up to 5 letters) is the measured data ranging
from -40.0 to 0.0 in dB. If measurement result is out of
range or Code Domain Power Pilot CH is not measured, the
value 99.9 is returned.
Code Domain Power Traffic (Fundamental) Measurement
TXANalyzer:RESults:CDPWr:TFUNDamental?
This query command returns the Code Domain Power Traffic
(Fundamental) CH Measurement result values, <int 1>, <int 2>,
and <real 3>. Each value is separated by a comma.
• <int 1> (1 digit) is 0 if measurement result is within the
measurement range, or 1 if measurement result is out of
range or Code Domain Power Traffic (Fundamental) CH is
not measured.
• <int 2> (1 digit) is the test result for the 0 (pass) or 1 (fail).
(Judgment of limit value is not executed.)