Agilent Technologies N9360A Programming Manual page 146

Multi ue tester
Hide thumbs Also See for N9360A:
Table of Contents

Advertisement

5
Programming Command Reference
5-96
• <real 3> (up to 5 letters) is the measured data ranging
from 0.900 to 0.999. If measurement result is out of range
or Rho is not measured, the value 9.999 is returned.
Multi-code Rho Measurement
TESTs:AUTO:RESults:MRHO1?
TESTs:AUTO:RESults:MRHO2?
TESTs:AUTO:RESults:MRHO3?
TESTs:AUTO:RESults:MRHO4?
TESTs:AUTO:RESults:MRHO5?
TESTs:AUTO:RESults:MRHO6?
These query commands are valid in the following case:
When CDMA mode is set to EDVO, or When CDMA mode is set to
MC1X and F3R3, F4R3, or F5R4 is set by the Radio Configuration
command. They return the Multi-code Rho Measurement result
values, <int 1>, <int 2> and <real 3> for each channel.
• <int 1> (1 digit) is 0 if measurement result is within the
measurement range, or 1 if measurement result is out of
range or Multi-code Rho is not measured.
• <int 2> (1 digit) is the test result for the 0 (pass) or 1 (fail).
• <real 3> (up to 5 letters) is the measured data ranging
from 0.900 to 0.999. If measurement result is out of range
or Multi-code Rho is not measured, the value 9.999 is
returned.
Time Offset Measurement
TESTs:AUTO:RESults:TOFFset1?
TESTs:AUTO:RESults:TOFFset2?
TESTs:AUTO:RESults:TOFFset3?
TESTs:AUTO:RESults:TOFFset4?
TESTs:AUTO:RESults:TOFFset5?
TESTs:AUTO:RESults:TOFFset6?
N9360A Multi UE Tester CDMA2000 Programming Manual