R&S FSV-K7 Operating Manual page 96

Analog demodulation and fm stereo
Table of Contents

Advertisement

®
R&S
FSV-K7/K7S
Note that you should perform a complete measurement with synchronization to the end
of the measurement. This is only possible for single sweeps.
Suffix:
<n>
<m>
Parameters:
<Range>
Example:
Manual operation:
CALCulate<n>:MARKer<m>:LINK <DisplayType>
Links the markers in all displays of the specified type.
Suffix:
<n>
<m>
Parameters:
<DisplayType>
Example:
Mode:
Manual operation:
CALCulate<n>:THReshold <Threshold>
This command defines a threshold value for the marker peak search.
Operating Manual 1176.7561.02 ─ 06
Analog Demodulation Option R&S FSV–K7
Remote Commands of the Analog Demodulation (R&S FSV–K7)
.
Selects the measurement window.
Selects the marker.
<numeric_value>
INIT:CONT OFF
Switches to single sweep mode
CALC:MARK:FUNC:ZOOM 1kHz;*WAI
Activates zooming and waits for its end.
See
"Marker Zoom (span > 0)"
.
1...4
window
1...16
marker
TIME | SPECtrum | BOTH | NONE
TIME
Links the markers in all time domain diagrams
SPECtrum
Links the markers in all AF Spectrum displays
BOTH
Links the markers both in the time domain diagrams and in the
AF Spectrum displays
NONE
Markers are not linked.
*RST:
NONE
CALC1:MARK1:LINK TIME
Links the marker 1 in all time domain diagrams in screen A.
ADEMOD
See
"Link Time Marker"
See
"Link AF Spectrum Marker"
on page 71
on page 71
on page 71
96

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fsv-k7s

Table of Contents