R&S FSV-K7 Operating Manual page 144

Analog demodulation and fm stereo
Table of Contents

Advertisement

®
R&S
FSV-K7/K7S
Manual operation:
[SENSe:]ADEMod<n>:ZOOM:STARt <Time>
The command selects the start time for the display of individual measured values of
the analog demodulation in the specified window. The maximum possible value
depends on the measurement time, which is set in the instrument and can be queried
with the
If the zoom function is enabled, the default number of sweep points in "Spectrum"
mode of the result memory are displayed from the specified start time with
]ADEMod<n>:ZOOM:STARt
Suffix:
<n>
Parameters:
<Time>
Example:
Mode:
Manual operation:
[SENSe:]ADEMod<n>:ZOOM:LENGth <Length>
The command allows you to define the length of the zoom area for the analog-
demodulated measurement data in the specified window manually. If the length is
defined manually using this command, the zoom mode is also set to manual.
Suffix:
<n>
Parameters:
<Length>
Example:
Mode:
Manual operation:
Operating Manual 1176.7561.02 ─ 06
Remote Commands of the Analog Demodulation (R&S FSV–K7)
See
"State On / Off"
[SENSe:]ADEMod:MTIMe
on page 144.
.
1...4
window
*RST:
0 s to measurement time – (default number of sweep points in
"Spectrum" mode – 1 * 1/sample rate)
ADEM:ZOOM:STAT ON (see
STATe]
Switches on the zoom function
ADEM:ZOOM:STAR 500us
Sets the starting point of the display to 500 µs.
ADEMOD
See
"Start"
.
1...4
window
*RST:
Length of the zoom area in seconds.
ADEM:ZOOM:LENG 2s
Zoom mode is set to manual and the zoom length to 2 seconds.
ADEMOD
See
"Length Manual"
Analog Demodulation Option R&S FSV–K7
on page 36
on page 133 command.
0 s
[SENSe:]ADEMod<n>:ZOOM[:
on page 143)
on page 36
sweep time
on page 36
[SENSe:
144

Advertisement

Table of Contents