Continuous Sweep
Sets the continuous sweep mode: the sweep takes place continuously according to the
trigger settings. This is the default setting.
The trace averaging is determined by the sweep count value (see the "Sweep Count"
softkey,
Remote command:
INIT:CONT ON, see
Single Sweep
Sets the single sweep mode: after triggering, starts the number of sweeps that are
defined by using the
number of sweeps has been performed.
Remote command:
INIT:CONT OFF, see
Continue Single Sweep
Repeats the number of sweeps set by using the
the trace of the last measurement.
This is particularly of interest when using the trace configurations "Average" or "Max
Hold" to take previously recorded measurements into account for averaging/maximum
search.
Remote command:
INITiate<n>:CONMeas
Meas Time
Opens an editor for entering the measurement time of the analog demodulation. For
details on the measurement time values refer to
surement Time and Trigger
Note: For FM Stereo measurements (option K7S), the minimum measurement time is
2 ms.
Remote command:
[SENSe:]ADEMod:MTIMe
Sweep Count
Opens an edit dialog box to enter the number of sweeps to be performed in the single
sweep mode. Values from 0 to 32767 are allowed. If the values 0 or 1 are set, one
sweep is performed. The sweep count is applied to all the traces in a diagram.
If the trace configurations "Average", "Max Hold" or "Min Hold" are set, the sweep
count value also determines the number of averaging or maximum search procedures.
In continuous sweep mode, if sweep count = 0 (default), averaging is performed over
10 sweeps. For sweep count =1, no averaging, maxhold or minhold operations are per-
formed.
Remote command:
[SENSe:]SWEep:COUNt
Operating Manual 1176.7561.02 ─ 06
"Sweep Count"
on page 56).
INITiate<n>:CONTinuous
Sweep Count
INITiate<n>:CONTinuous
on page 185
Offset",
on page 133
on page 155
Analog Demodulation Option R&S FSV–K7
Softkeys of the Analog Demodulation option
on page 186
softkey. The measurement stops after the defined
on page 186
Sweep Count
Chapter 2.1.3, "Sample Rate, Mea-
on page 13.
softkey, without deleting