Detector Overview - R&S FSV-K7 Operating Manual

Analog demodulation and fm stereo
Table of Contents

Advertisement

®
R&S
FSV-K7/K7S
Remote command:
DISP:TRAC:MODE VIEW, see
on page 110
Blank
Hides the selected trace.
Remote command:
DISP:TRAC OFF, see

2.1.8 Detector Overview

The measurement detector for the individual display modes can be selected directly by
the user or set automatically by the R&S FSVA/FSV. The detector activated for the
specific trace is indicated in the corresponding trace display field by an abbreviation.
The detectors of the R&S FSVA/FSV are implemented as pure digital devices. They
collect signal power data within each measured point during a sweep. The default
number of sweep points is 691. The following detectors are available:
Table 2-3: Detector types
Detector
Auto Peak
Positive Peak
Negative Peak (min peak)
RMS
Average
Sample
The result obtained from the selected detector within a measurement point is displayed
as the power value at this measurement point.
All detectors work in parallel in the background, which means that the measurement
speed is independent of the detector combination used for different traces.
Operating Manual 1176.7561.02 ─ 06
Analog Demodulation Option R&S FSV–K7
Instrument Functions Analog Demodulation
DISPlay[:WINDow<n>]:TRACe<t>:MODE
DISPlay[:WINDow<n>]:TRACe<t>[:STATe]
Indicator
Function
Ap
Determines the maximum and the minimum value
within a measurement point (not available for SEM)
Pk
Determines the maximum value within a measure-
ment point
Mi
Determines the minimum value within a measure-
ment point
Rm
Determines the root mean square power within a
measurement point
Av
Determines the linear average power within a mea-
surement point
Sa
Selects the last value within a measurement point
on page 111