Note that continuous averaging is performed after the indicated number has been
reached in continuous sweep mode.
In single sweep mode, the sweep is stopped as soon as the indicated number of mea-
surements (sweeps) is reached. Synchronization to the end of the indicated number of
measurements is only possible in single sweep mode.
This command has the same effect as the [SENSe<source>:]SWEep:COUNt com-
mand. In both cases, the number of measurements is defined whether the average cal-
culation is active or not.
The number of measurements applies to all traces in the window.
Suffix:
<n>
Parameters:
<NoMeasurements> 0 to 32767
Example:
[SENSe:]AVERage<n>:TYPE <FunctionType>
This command selects the type of average function.
Suffix:
<n>
Parameters:
<FunctionType>
Example:
Operating Manual 1176.7561.02 ─ 06
Analog Demodulation Option R&S FSV–K7
Remote Commands of the Analog Demodulation (R&S FSV–K7)
.
Selects the measurement window.
*RST:
0
SWE:CONT OFF
Switching to single sweep mode.
AVER:COUN 16
Sets the number of measurements to 16.
AVER:STAT ON
Switches on the calculation of average.
INIT;*WAI
Starts the measurement and waits for the end of the 16 sweeps.
.
Selects the measurement window.
VIDeo | LINear | POWer
VIDeo
The logarithmic power values are averaged.
LINear
The power values are averaged before they are converted to
logarithmic values.
POWer
The power level values are converted into unit Watt prior to
averaging. After the averaging, the data is converted back into
its original unit.
*RST:
VIDeo
AVER:TYPE LIN
Switches to linear average calculation.