R&S FSV-K7 Operating Manual page 201

Analog demodulation and fm stereo
Table of Contents

Advertisement

®
R&S
FSV-K7/K7S
25 µs:
50 µs:
75 µs:
750 µs:
The following table shows the required demodulation bandwidth for an error less than
0.5 dB up to a maximum AF frequency.
deemphasis
max. AF frequency
required demodulation bandwidth
For higher AF frequencies the demodulation bandwidth must be increased.
Remote command:
[SENSe:]FILTer<n>:DEMPhasis[:STATe]
[SENSe:]FILTer<n>:DEMPhasis:TCONstant
SFM:
[SENSe:]SFM:<ChannelType>:FILTer:DEMPhasis:STATe
[SENSe:]SFM:<ChannelType>:FILTer:DEMPhasis:TCONstant
Coupled ← AF Filter ← Left
All channels for which this option is enabled are configured identically, i.e. the channel
settings are coupled. If you change the settings for one coupled channel, the settings
are changed for all other coupled channels, as well. The settings are taken from the
first channel for which coupling is enabled.
Remote command:
[SENSe:]SFM:<ChannelType>:RSUMmary:COUPling
All AF Filter Off ← AF Filter ← Left
Disables all specified AF Filters.
Remote command:
[SENSe:]FILTer<n>:AOFF
AF Range ← Left
Opens a submenu to define the diagram scaling for AF displays.
Dev per Division ← AF Range ← Left
Opens an edit dialog box to set the modulation depth or the phase deviation (Analog
Demodulation only), or frequency deviation per division:
AM display:
FM display:
PM display:
The softkey is not available if logarithmic display is set ( "Deviation Lin/Log" softkey).
Operating Manual 1176.7561.02 ─ 06
Instrument Functions FM Stereo (R&S FSV–K7S)
25 kHz ≤ demodulation bandwidth ≤ 40 MHz
6.4 kHz ≤ demodulation bandwidth ≤ 18 MHz
6.4 kHz ≤ demodulation bandwidth ≤ 18 MHz
800 Hz ≤ demodulation bandwidth ≤ 3 MHz
25 µs
25 kHz
≥ 200 kHz
on page 148
0.0001 % to 1000 %
1 Hz/div to 100 MHz/div
0.0001 rad/div to 1000 rad/div
FM Stereo Option R&S FSV–K7S
50 µs
75 µs
12 kHz
8 kHz
≥ 100 kHz
≥ 50 kHz
on page 150
on page 151
on page 239
on page 242
750 µs
800 Hz
≥ 6.4 kHz
on page 239