CALCulate<n>:MARKer:FUNCtion:ADEMod:CARRier[:RESult<t>]?
This command queries the carrier power.
With RF Power result display, the carrier power is determined from trace 1 to 6 indica-
ted in the suffix. With all other result displays, the carrier power is determined from the
current trace data (CLR/WRITE trace).
Suffix:
<n>
<t>
Example:
Usage:
Mode:
CALCulate<n>:MARKer:FUNCtion:ADEMod:FERRor[:RESult<t>]?
This command queries the frequency error with FM and PM demodulation. The fre-
quency error is determined from the current measurement data (CLR/WRITE trace).
The offset thus determined differs from that calculated in the
OFFSet
the modulation is removed by means of low pass filtering, producing results that are
different from those obtained by averaging with the SENSe command.
Suffix:
<n>
<t>
Operating Manual 1176.7561.02 ─ 06
Remote Commands of the Analog Demodulation (R&S FSV–K7)
Query only
ADEMOD
.
1...4
irrelevant
1...6
irrelevant
ADEM ON (see
Switches on analog demodulator
CALC:FEED 'XTIM:RFP', see
on page 90
Switches on RF power result display
CALC:MARK:FUNC:ADEM:CARR?
Queries the carrier power
Query only
ADEMOD
on page 132 command since, for determination of the frequency deviation,
.
1...4
irrelevant
1...6
irrelevant
Analog Demodulation Option R&S FSV–K7
[SENSe:]ADEMod[:STATe]
CALCulate<n>:FEED
on page 121)
[SENSe:]ADEMod:FM:
87