R&S FSV-K7 Operating Manual page 106

Analog demodulation and fm stereo
Table of Contents

Advertisement

®
R&S
FSV-K7/K7S
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT <Limit>
This command sets the left limit of the marker search range.
If the power measurement in zero span is active, this command limits the evaluation
range to the trace.
Note: The function is only available if the search limit for marker and delta marker is
switched on (see CALCulate<n>:MARKer<m>:X:SLIMits[:STATe]).
Suffix:
<n>
<m>
Parameters:
<Limit>
Example:
Manual operation:
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHT <Limit>
This command sets the right limit of the marker search range.
If the power measurement in zero span is active, this command limits the evaluation
range to the trace.
Note: The function is only available if the search limit for marker and delta marker is
switched on ( CALCulate<n>:MARKer<m>:X:SLIMits[:STATe]).
Suffix:
<n>
<m>
Parameters:
<Limit>
Example:
Operating Manual 1176.7561.02 ─ 06
Analog Demodulation Option R&S FSV–K7
Remote Commands of the Analog Demodulation (R&S FSV–K7)
.
Selects the measurement window.
irrelevant
The value range depends on the span or sweep time.
The unit is Hz for frequency domain measurements and s for
time domain measurements.
Range:
0 to MAX
*RST:
left diagram border
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:LEFT 10MHz
Sets the left limit of the search range to 10 MHz.
See
"Left Limit"
on page 74
.
Selects the measurement window.
irrelevant
The value range depends on the span or sweep time.
The unit is Hz for frequency domain measurements and s for
time domain measurements.
Range:
0 to MAX
*RST:
left diagram border
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:RIGH 20MHz
Sets the right limit of the search range to 20 MHz.
106

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fsv-k7s

Table of Contents