R&S FSV-K7 Operating Manual page 166

Analog demodulation and fm stereo
Table of Contents

Advertisement

®
R&S
FSV-K7/K7S
Example:
Manual operation:
TRIGger<n>[:SEQuence]:LEVel:RFPower <TriggerLevel>
This command defines the power level at the third intermediate frequency that must be
exceeded to cause a trigger event. Note that any RF attenuation or preamplification is
considered when the trigger level is analyzed. If defined, a reference level offset is also
considered.
Suffix:
<n>
Parameters:
<TriggerLevel>
Example:
TRIGger<n>[:SEQuence]:LEVel:PM <Level>
The command sets the level when PM-modulated signals are used as trigger source.
For triggering with AF, AM, AMRelative, FM, and PM trigger sources to be successful,
the measurement time must cover at least 5 periods of the audio signal.
Suffix:
n
Parameters:
<Level>
Example:
Mode:
TRIGger<n>[:SEQuence]:TIME:RINTerval <Interval>
This command sets the repetition interval for the time trigger source.
Suffix:
<n>
Parameters:
<Interval>
Example:
Operating Manual 1176.7561.02 ─ 06
Analog Demodulation Option R&S FSV–K7
Remote Commands of the Analog Demodulation (R&S FSV–K7)
TRIG:LEV:IFP -30DBM
See
"Trigger Level"
on page 66
.
irrelevant
*RST:
-20 dBm
TRIG:LEV:RFP -30dBm
.
irrelevant
Range:
-1000 to +1000
*RST:
0 RAD
Default unit: RAD | DEG
TRIG:LEV:PM 1.2 RAD
Sets the PM trigger threshold to 1.2 rad
ADEMOD
.
irrelevant
2.0 ms to 5000
*RST:
1.0
TRIG:SOUR TIME
Selects the time trigger input for triggering.
TRIG:TIME:RINT 50
The sweep starts every 50 s.
166

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fsv-k7s

Table of Contents