CALCulate<n>:DELTamarker<m>:Y?
This command queries the measured value of a delta marker. The corresponding delta
marker is activated, if necessary. The output is always a relative value referred to
marker 1 or to the reference position (reference fixed active).
To obtain a correct query result, a complete sweep with synchronization to the sweep
end must be performed between the activation of the delta marker and the query of the
y value. This is only possible in single sweep mode.
Depending on the unit defined with CALC:UNIT:POW or on the activated measuring
functions, the query result is output in the units below:
Table 2-6: Analog demodulation measurements
Parameter, measuring function or result display
AM result display (R&S FSV–K7)
FM result display (R&S FSV–K7)
PM result display (R&S FSV–K7)
RF result display (R&S FSV–K7)
Suffix:
<n>
<m>
Example:
Usage:
Manual operation:
2.3.10.2
INPut subsystem
INPut:ATTenuation
INPut:ATTenuation:AUTO
INPut:COUPling
INPut:DIQ:CDEVice
INPut:DIQ:RANGe:COUPling
Operating Manual 1176.7561.02 ─ 06
Remote Commands of the Analog Demodulation (R&S FSV–K7)
.
Selects the measurement window.
Selects the marker.
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
CALC:DELT2 ON
Switches on delta marker 2.
CALC:DELT2:Y?
Outputs measurement value of delta marker 2.
Query only
See
"Marker 1 / Marker 2 / Marker 3 / ... Marker 16,/ Marker
Norm/Delta"
on page 68
......................................................................................................... 177
............................................................................................... 177
.............................................................................................................178
........................................................................................................178
.......................................................................................... 179
Analog Demodulation Option R&S FSV–K7
Output unit
% (lin)
dB (log)
Hz (lin)
dB (log)
rad | deg (lin)
dB (log)
dB (Range Log or Range Linear %)
% (Range Linear %)