In single sweep mode, you can synchronize to the end of the measurement with *OPC,
*OPC? or *WAI. In continuous sweep mode, synchronization to the end of the mea-
surement is not possible. Thus, it is not recommended that you use continuous sweep
mode in remote control, as results like trace data or markers are only valid after a sin-
gle sweep end synchronization.
Suffix:
<n>
Example:
Mode:
SYSTem:DISPlay:UPDate <State>
In remote control mode, this command switches on or off the instrument display. If
switched on, only the diagrams, traces and display fields are displayed and updated.
The best performance is obtained if the display output is switched off during remote
control.
Parameters:
<State>
Example:
Operating Manual 1176.7561.02 ─ 06
Analog Demodulation Option R&S FSV–K7
Remote Commands of the Analog Demodulation (R&S FSV–K7)
.
irrelevant
INIT:CONT OFF
Switches to single sweep mode.
DISP:WIND:TRAC:MODE AVER
Switches on trace averaging.
SWE:COUN 20
Setting the sweep counter to 20 sweeps.
INIT;*WAI
Starts the measurement and waits for the end of the 20 sweeps.
all
ON | OFF
*RST:
OFF
SYST:DISP:UPD ON
187