R&S FSV-K7 Operating Manual page 58

Analog demodulation and fm stereo
Table of Contents

Advertisement

®
R&S
FSV-K7/K7S
The default setting is trace 1 in the overwrite mode (see
other traces are switched off (see
"Trace Mode
Tip: To configure several traces in one step, press the
trace configuration dialog. See also
Remote command:
Selected via numeric suffix of:TRACe<1...6> commands
Clear Write ← Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
Overwrite mode: the trace is overwritten by each sweep. This is the default setting.
All available detectors can be selected.
Remote command:
DISP:TRAC:MODE WRIT, see
on page 110
Max Hold ← Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
The maximum value is determined over several sweeps and displayed. The
R&S FSVA/FSV saves the sweep result in the trace memory only if the new value is
greater than the previous one.
The detector is automatically set to "Positive Peak".
This mode is especially useful with modulated or pulsed signals. The signal spectrum
is filled up upon each sweep until all signal components are detected in a kind of enve-
lope.
This mode is not available for statistics measurements.
Remote command:
DISP:TRAC:MODE MAXH, see
on page 110
Min Hold ← Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
The minimum value is determined from several measurements and displayed. The
R&S FSVA/FSV saves the smallest of the previously stored/currently measured values
in the trace memory.
The detector is automatically set to "Negative Peak".
This mode is useful e.g. for making an unmodulated carrier in a composite signal visi-
ble. Noise, interference signals or modulated signals are suppressed whereas a CW
signal is recognized by its constant level.
This mode is not available for statistics measurements.
Remote command:
DISP:TRAC:MODE MINH, see
on page 110
Average ← Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
The average is formed over several sweeps. The
of averaging procedures.
All available detectors can be selected. If the detector is automatically selected, the
sample detector is used (see
Operating Manual 1176.7561.02 ─ 06
"Blank"
Overview",
on page 17.
DISPlay[:WINDow<n>]:TRACe<t>:MODE
DISPlay[:WINDow<n>]:TRACe<t>:MODE
DISPlay[:WINDow<n>]:TRACe<t>:MODE
Chapter 2.1.8, "Detector
Analog Demodulation Option R&S FSV–K7
Softkeys of the Analog Demodulation option
"Clear Write"
on page 19). For details see
Trace Wizard
Chapter 2.1.6, "Configuring
Sweep Count
Overview",
on page 17), the
Chapter 2.1.7,
softkey to open a
Traces",
on page 16.
determines the number
on page 19).
58