[SENSe:]ADEMod:BANDwidth|BWIDth:DEModulation <Bandwidth>
This command defines the demodulation bandwidth used for analog demodulation.
The required sampling rate is automatically set depending on the selected demodula-
tion bandwidth. The available demodulation bandwidths are determined by the existing
sampling rates. For details on the relation between demodulation bandwidth and sam-
pling rate refer to
set",
Parameters:
<Bandwidth>
Example:
Mode:
[SENSe:]ADEMod:BANDwidth|BWIDth:DEModulation:TYPE <FilterType>
This command defines the type of demodulation filter to be used.
Parameters:
<FilterType>
Example:
Mode:
[SENSe:]ADEMod:FM[:TDOMain][:TYPE] <Type>
This command selects the result types to be measured simultaneously by FM demodu-
lation.
Parameters:
<Type>
Operating Manual 1176.7561.02 ─ 06
Remote Commands of the Analog Demodulation (R&S FSV–K7)
Chapter 2.1.3, "Sample Rate, Measurement Time and Trigger Off-
on page 13.
*RST:
For details on the correlation of bandwidth and sample rate refer
to chapter "Instrument Functions", section "Analog Demodula-
tion (Option K7)" –
Time and Trigger
ADEM:BAND:DEM 1MHz
Sets the demodulation bandwidth to 1 MHz.
ADEMOD
FLAT
Standard flat demodulation filter
GAUSs
Gaussian filter for optimized settling behaviour
*RST:
BAND:DEM:TYPE GAUS
Selects the Gaussian filter.
ADEMOD
*RST:
<result type 1|2|3|4|5|6>: WRITe, AVERage, MAXHold, MINHold,
VIEW, OFF; for details see
Types",
Analog Demodulation Option R&S FSV–K7
5 MHz
Chapter 2.1.3, "Sample Rate, Measurement
Offset",
on page 13.