[Sense:]Ademod:fm[:Tdomain][:Type] - R&S FSV-K7 Operating Manual

Analog demodulation and fm stereo
Table of Contents

Advertisement

®
R&S
FSV-K7/K7S
Example:
Mode:
[SENSe:]ADEMod:FM[:TDOMain]:RESult <Type>
This command reads the result data obtained by analog demodulation for the specified
result type. The data format of the output data block is defined by the FORMat com-
mand.
Return values:
<Type>
Return values
Example:
Operating Manual 1176.7561.02 ─ 06
Analog Demodulation Option R&S FSV–K7
Remote Commands of the Analog Demodulation (R&S FSV–K7)

[SENSe:]ADEMod:FM[:TDOMain][:TYPE]

"Creates average, max hold and min hold values simultaneously
DEM:FM WRIT,OFF,OFF
Only creates the current measurement values
ADEM:FM OFF,OFF,OFF
Switches analog demodulator off
ADEMOD
<result type>: WRITe, AVERage, MAXHold, MINHold; for details
see
Chapter 2.3.6.1, "Trace Mode Result
The result type indicated must be one of those configured by
[SENSe:]ADEMod:FM[:TDOMain][:TYPE]
Otherwise a query error will be generated.
ASCII Format (FORMat ASCII) or Binary Format (FORMat
REAL,32); for details see
Returned Values: ASCII Format and Binary
on page 118 .
Default unit: Hz
ADEM:SET 8MHz,32000,EXT,POS,-500,30, see
on page 138
]ADEMod:SET
Sets up demodulator parameters
ADEM:FM AVER,MAXH,MINH, see
on page 129
TDOMain][:TYPE]
Selects FM results to be measured
ADEM:AM WRIT,OFF,OFF
Selects AM results to be measured
ADEM ON, see
[SENSe:]ADEMod[:STATe]
Switches on demodulator
INIT; WAI
Starts measurement and waits for sync
FORM ASC, see
FORMat[:DATA]
Selects output format
ADEM:FM:RES? AVER
Reads FM average results
ADEM:FM:RES? MAXH
Reads FM max hold results
ADEM:FM:RES? MINH
Reads FM min hold results
ADEM:AM:RES? WRIT
Reads current AM results
on page 129
Types",
on page 129.
Chapter 2.3.6.2, "Formats for
Format",
[SENSe:]ADEMod:FM[:
on page 121
on page 184
on page 118.
[SENSe:
130

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fsv-k7s