R&S FSV-K7 Operating Manual page 146

Analog demodulation and fm stereo
Table of Contents

Advertisement

®
R&S
FSV-K7/K7S
Parameters:
<Threshold>
Example:
Example:
Manual operation:
[SENSe:]ADJust:CONFiguration:HYSTeresis:UPPer <Threshold>
This command defines an upper threshold the signal must exceed before the reference
level is automatically adjusted when the "Auto Level" function is performed.
For more information see [SENSe:]ADJust:LEVel).
Parameters:
<Threshold>
Example:
Manual operation:
[SENSe:]ADJust:CONFigure:LEVel:DURation <Duration>
This command defines the duration of the level measurement used to determine the
optimal reference level automatically (for SENS:ADJ:LEV ON).
Parameters:
<Duration>
Example:
Manual operation:
[SENSe:]ADJust:FREQuency
This command defines the center frequency and the reference level automatically by
determining the highest level in the frequency span.
Example:
Manual operation:
Operating Manual 1176.7561.02 ─ 06
Analog Demodulation Option R&S FSV–K7
Remote Commands of the Analog Demodulation (R&S FSV–K7)
Range:
0 to 200
*RST:
+1 dB
Default unit: dB
SENS:ADJ:CONF:HYST:LOW 2
For an input signal level of currently 20 dBm, the reference level
will only be adjusted when the signal level falls below 18 dBm.
See
"Lower Level Hysteresis"
Range:
0 to 200
*RST:
+1 dB
Default unit: dB
SENS:ADJ:CONF:HYST:UPP 2
For an input signal level of currently 20 dBm, the reference level
will only be adjusted when the signal level rises above 22 dBm.
See
"Upper Level Hysteresis"
<numeric value> in seconds
Range:
0.001 to 16000.0
*RST:
0.001
Default unit: s
ADJ:CONF:LEV:DUR:5
See
"Meas Time Manual"
ADJ:FREQ
See
"Auto Freq"
on page 50
on page 51
on page 50
on page 50