R&S
FSV-K7/K7S
In single sweep mode, you can synchronize to the end of the measurement with *OPC,
*OPC? or *WAI. In continuous sweep mode, synchronization to the end of the mea-
surement is not possible. Thus, it is not recommended that you use continuous sweep
mode in remote control, as results like trace data or markers are only valid after a sin-
gle sweep end synchronization.
Suffix:
<n>
Example:
Manual operation:
INITiate<n>:CONTinuous <State>
This command determines whether the trigger system is continuously initiated (contin-
uous) or performs single measurements (single).
The sweep is started immediately.
Suffix:
<n>
Parameters:
<State>
Example:
Mode:
Manual operation:
INITiate<n>[:IMMediate]
The command initiates a new measurement sequence.
With sweep count > 0 or average count > 0, this means a restart of the indicated num-
ber of measurements. With trace functions MAXHold, MINHold and AVERage, the pre-
vious results are reset on restarting the measurement.
Operating Manual 1176.7561.02 ─ 06
Analog Demodulation Option R&S FSV–K7
Remote Commands of the Analog Demodulation (R&S FSV–K7)
.
irrelevant
INIT:CONT OFF
Switches to single sweep mode.
DISP:WIND:TRAC:MODE AVER
Switches on trace averaging.
SWE:COUN 20
Setting the sweep counter to 20 sweeps.
INIT;*WAI
Starts the measurement and waits for the end of the 20 sweeps.
INIT:CONM;*WAI
Continues the measurement (next 20 sequences) and waits for
the end.
See
"Continue Single Sweep"
.
irrelevant
ON | OFF
*RST:
ON
INIT:CONT OFF
Switches the sequence to single sweep.
INIT:CONT ON
Switches the sequence to continuous sweep.
all
See
"Continuous Sweep"
See
"Single Sweep"
on page 56
on page 56
on page 56
186