Mode:
Manual operation:
[SENSe:]ADEMod<n>:AF:STOP <Frequency>
This command sets the stop frequency for AF spectrum result display.
The stop frequency is limited to half the measurement bandwidth of analog demodula-
tion
([SENSe:]ADEMod:BANDwidth|BWIDth:DEModulation
Suffix:
<n>
Parameters:
<Frequency>
Example:
Mode:
Manual operation:
Operating Manual 1176.7561.02 ─ 06
Analog Demodulation Option R&S FSV–K7
Remote Commands of the Analog Demodulation (R&S FSV–K7)
ADEM ON, see
[SENSe:]ADEMod[:STATe]
Switches on the analog demodulator
CALC:FEED 'XTIM:FM:AFSP', see
on page 90
Switches on AF spectrum result display of FM
ADEM:BAND 5 MHz, see
BANDwidth|BWIDth:DEModulation
Sets the measurement bandwidth to 5 MHz
ADEM:AF:STAR 0 kHz
Sets the AF start frequency to 0 kHz
[SENSe:]ADEMod<n>:AF:STOP
Sets the AF stop frequency to 500 kHz
ADEMOD
See
"AF Start"
on page 42
.
1...4
irrelevant
*RST:
2.5 MHz
ADEM ON, see
[SENSe:]ADEMod[:STATe]
Switches on the analog demodulator
CALC:FEED 'XTIM:FM:AFSP', see
on page 90
Switches on AF spectrum result display of FM
ADEM:BAND 5 MHz, see
BANDwidth|BWIDth:DEModulation
Sets the measurement bandwidth to 5 MHz
[SENSe:]ADEMod<n>:AF:STARt
Sets the AF start frequency to 0 kHz
ADEM:AF:STOP 500 kHz
Sets the AF stop frequency to 500 kHz
ADEMOD
See
"AF Stop"
on page 42
on page 121
CALCulate<n>:FEED
[SENSe:]ADEMod:
on page 129
on page 124
on page 129).
on page 121
CALCulate<n>:FEED
[SENSe:]ADEMod:
on page 129
on page 123
124

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fsv-k7s

Table of Contents