R&S FSV-K7 Operating Manual page 14

Analog demodulation and fm stereo
Table of Contents

Advertisement

®
R&S
FSV-K7/K7S
Option K7S
The K7S option always uses the demodulation bandwidth 400 kHz.
Table 2-1: Sample Rate, Measurement Time and Trigger Offset using a flat demodulation filter
Demod. band-
width
160 MHz
80 MHz
40 MHz
28 MHz
18 MHz
10 MHz
8 MHz
5 MHz
3 MHz
1.6 MHz
800 kHz
400 kHz
200 kHz
100 kHz
50 kHz
25 kHz
12.5 kHz
6.4 kHz
3.2 kHz
1.6 kHz
800 Hz
400 Hz
200 Hz
100 Hz
1)
only available with option R&S FSV-B160
2)
only available with option R&S FSV-B70/R&S FSVA-B40
Operating Manual 1176.7561.02 ─ 06
Sample rate
200 MHz
1)
2)
128 MHz
2)
64 MHz
64 MHz
32 MHz
32 MHz
16 MHz
8 MHz
4 MHz
2 MHz
1 MHz
500 kHz
250 kHz
125 kHz
62.5 kHz
31.25 kHz
15.625 kHz
7.8125 kHz
3.90625 kHz
1.953125 kHz
976.5625 Hz
488.28125 Hz
244.140625 Hz
122.0703125
Hz
Analog Demodulation Option R&S FSV–K7
Instrument Functions Analog Demodulation
Measurement time
Min.
Max.
5 ns
8 ms
7.8125 ns
12.5 ms
15.625 ns
25 ms
15.625 ns
25 ms
31.25 ns
50 ms
31.25 ns
50 ms
62.5 ns
100 ms
125 ns
200 ms
250 ns
400 ms
500 ns
800 ms
1 µs
1.6 s
2 µs
3.2 s
4 µs
6.4 s
8 µs
12.8 s
16 µs
25.6 s
32 µs
51.2 s
64 µs
102.4 s
128 µs
204.8 s
256 µs
409.6 s
512 µs
819.2 s
1.024 ms
1638.4 s
2.048 ms
3276.8 s
4.096 ms
6553.6 s
8.192 ms
13107.2 s
Trigger offset
Min.
Max.
-8 ms
1.0486 s
-12.5 ms
1.6384 s
-25 ms
3.2768 s
-25 ms
3.2768 s
-50 ms
6.5536 s
-50 ms
6.5536 s
-100 ms
13.1072 s
-200 ms
26.2144 s
-400 ms
52.4288 s
-800 ms
104.8576 s
-1.6 s
209.7152 s
-3.2 s
419.4304 s
-6.4 s
838.8608 s
-12.8 s
1677.7216 s
-25.6 s
3355.4432 s
-51.2 s
6710.8864 s
-102.4 s
13421.7728 s
-204.8 s
26843.5456 s
-409.6 s
53687.0912 s
-819.2 s
107374.1824 s
-1638.4 s
214748.3648 s
-3276.8 s
429496.7296 s
-6553.6 s
858993.4592 s
-13107.2 s
1717986.918 s
14