Options and Accessories
How to Order Options

Rack Mount Kit With Handles (Option 1CP)

Option 1CP provides the parts necessary to mount the analyzer in an
HP/Agilent System II cabinet or in a standard 19 inch (482.6 mm)
equipment rack. It includes front handles and rack slides for added
convenience.

High Stability Frequency Reference (Option 1D5)

Option 1D5 improves the frequency reference accuracy. The analyzer's
synthesizer is phase locked to an oven controlled crystal oscillator
(OCXO), instead of the standard VCXO.
Option 1D5 is only available at the time of purchase, except for the
Agilent E4401B spectrum analyzer which can be retrofitted.

Time-Gated Spectrum Analysis (Option 1D6)

Option 1D6 allows you to selectively measure the spectrum of signals
that may overlap in the frequency domain, but are separated in the
time domain. By adjusting a time gate based on a user-supplied trigger
signal, you can significantly increase the diagnostic capability of your
spectrum analyzer for time-interleaved signals.
50 Ω to 75 Ω Matching Pad (Option 1D7)
This option provides a 50 Ω to 75 Ω matching pad with dc block to be
used on the analyzer input. The pad has a frequency range of 9 MHz to
2 GHz. It adapts your standard 50 Ω analyzer to be compatible with a
75 Ω system under test. Connector types are 50 Ω Type-N (m) to 75 Ω
BNC (f).
50 Ω Tracking Generator (Option 1DN)
Option 1DN provides a 9 kHz to 1.5 GHz built-in tracking generator for
the Agilent E4401B and the Agilent E4411B. Option 1DN provides a
9 kHz to 3 GHz built-in tracking generator for the Agilent E4402B,
E4403B, E4404B, E4405B, E4407B and E4408B. This source creates a
source-receiver combination that allows insertion-loss, frequency
response, and return-loss measurements. The tracking generator has a
wide distortion-free dynamic range, plus good sensitivity and
selectivity.
75 Ω Input Impedance (Option 1DP)
This option provides a 75 Ω input impedance instead of the standard
50 Ω impedance. Analyzers with this option use cables, circuit boards,
and front panels that are different from the standard units. Option 1DP
is only available on the Agilent E4401B and E4411B.
Option 1DP is not available after the purchase of your analyzer.
7-6
Chapter 7