Test Head Test; Dc Bias Test (Option 001 Only) - Agilent Technologies E4991A Service Manual

Rf impedance/material analyzer
Hide thumbs Also See for E4991A:
Table of Contents

Advertisement

Test Head Test

This test is to check if the test head works normally.
Required Equipment
Description
Calibration Kit
N(m)-N(m), 61cm Cable
Procedure
The following is the procedure to perform the Test Head Test.
Step 1. Make sure that noting is connected to the instrument.
Step 2. Open the External Test dialog box as described in "To Execute the External Test" on
page 189.
Step 3. Click
on External Test dialog box.
7.Test Head
Step 4. Connect RF OUT and PORT1 using N Cable. Then click
Step 5. Connect RF OUT and PORT2 using N Cable. Then click
Step 6. Connect the Test Head to the instrument, and connect the 50ohm Termination of the
16195B to DUT Port on the Test Head. Then click
Step 7. Click
OK & Next>>

DC Bias Test (Option 001 only)

This test is to check if the function of A2 DC Bias Board(Option 001) works normally.
Required Equipment
Description
Calibration Kit
Procedure
The following is the procedure to perform the DC Bias Test.
Step 1. Connect the Test Head to the instrument.
Step 2. Open the External Test dialog box as described in "To Execute the External Test" on
page 189.
Step 3. Click
on External Test dialog box.
8.DC Bias
Step 4. Connect the OPEN(0 S) Termination of the 16195B to DUT Port on the Test Head. Then
click
.
OK
Chapter 4
External Tests Failure Troubleshooting
Recommended Model
16195B (50 ohm)
11500B
to go to the next test.
Recommended Model
16195B (Open, Short, 50 ohm)
Troubleshooting
.
OK
.
OK
.
OK
193