Performance Tests

15. Source and Receiver Harmonics (Option 002 Only)

15. Source and Receiver Harmonics (Option 002 Only)
Perform this test to determine the 2nd and 3rd harmonics of the source and receiver.
Analyzer warmup time: 30 minutes
Specifications
a
Source Harmonics
2nd Harmonic 16 MHz–3 GHz
3rd Harmonic 16 MHz–2 GHz
b
Receiver Harmonics
2nd Harmonic
3rd Harmonic
Source and Receiver Harmonics
2nd Harmonic
3rd Harmonic
a. At +20 dBm output level (or +18 dBm for Option 006).
b. At 0 dBm input level.
Required Equipment
Description
2-way power splitter, 50Ω
Attenuator, 10 dB
Attenuator, 30 dB
RF cable set, 50Ω type-N
2-66
Power Level
< −25 dBc
<−25 dBc
< −15 dBc
<−30 dBc
<−15 dBc
<−30 dBc
HP/Agilent Model Number
11667A Option 001
8491A Option 010
8491A Option 030
11851B