Optimize Preselector Frequency
[:SENSe]:POWer[:RF]:PADJust <freq>
[:SENSe]:POWer[:RF]:PADJust?
This command allows user-defined adjustment of the preselector
frequency to optimize its response on the signal of interest.
Factory Preset
and *RST:
Range:
Default Unit:
Remarks:
Front Panel
Access:
Preselector Center
[:SENSe]:POWer[:RF]:PCENter
In internal mixing, this command centers the preselector filter at the
signal of interest. In external mixing, the external preselector filter is
adjusted to the peak of the filter response to maximize the amplitude at
the active marker frequency. This command has no effect if it is
activated in non-preselected bands.
This command is available only on Agilent ESA models E4404B,
NOTE
E4405B, E4407B, and E4408B. This command has no effect with
markers set to less than 3 GHz.
Remarks:
Front Panel
Access:

[:SENSe]:SIDentify Subsection

Set Mixer Signal Identification Mode
[:SENSe]:SIDentify:MODE ISUPpress|ISHift
[:SENSe]:SIDentify:MODE?
Chapter 5
0 Hz
–250 MHz to 250 MHz
None. Use the MHz terminator in order for this
command to work.
This command is available only on Agilent ESA models
E4404B, E4405B, E4407B, and E4408B. Use this
command for signals close to the noise level, multiple
signals close together, or for other conditions when the
preselector is not tuned to the frequency of interest.
AMPLITUDE/Y Scale, Presel Adjust
A peak search will be done if no marker is on.
AMPLITUDE/Y Scale, Presel Center
Language Reference
SENSe Subsystem
5-103