Test Port Input

Table 15-10
Test Port Input Levels
Description
Maximum Test Port Input Level
300 kHz to 8.5 GHz
Damage Level
300 kHz to 8.5 GHz
*1
Crosstalk
3 MHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 7.5 GHz
7.5 GHz to 8.5 GHz
*1.Response calibration not omitted.
Table 15-11
Test Port Input (Trace Noise)
Description
*1
Trace Noise
Magnitude
300 kHz to 3 MHz
3 MHz to 4.2 GHz
4.2 GHz to 7.5 GHz
7.5 GHz to 8.5 GHz
Chapter 15
Specifications and Supplemental Information
Specification
+0 dBm max.
-120 dB
-110 dB
-100 dB
-90 dB
Specification
1 mdB rms
(at IFBW = 3 kHz)
3 mdB rms
(at IFBW = 3 kHz)
5 mdB rms
(at IFBW = 3 kHz)
Test Port Input
Supplemental Information
+20 dBm, ±25 VDC, typical
Supplemental Information
5 mdB rms
(at IFBW = 3 kHz, typical)
8 mdB rms
(at IFBW = 3 kHz, high
temperature mode: ON, typical)
4 mdB rms
(at IFBW = 3 kHz, high
temperature mode: ON, typical)
6 mdB rms
(at IFBW = 3 kHz, high
temperature mode: ON, typical)
8 mdB rms
(at IFBW = 3 kHz, high
temperature mode: ON, typical)