Increasing Dynamic Range; Increase The Test Port Input Power; Reduce The Receiver Noise Floor; Reduce The Receiver Crosstalk - Agilent Technologies 8719ES User Manual

Network analyzer
Hide thumbs Also See for 8719ES:
Table of Contents

Advertisement

Increasing Dynamic Range

Dynamic range is the difference between the analyzer's maximum allowable input level
and minimum measurable power. For a measurement to be valid, input signals must be
within these boundaries. The dynamic range is affected by these factors:
• test port input power
• test port noise floor
• receiver crosstalk

Increase the Test Port Input Power

You can increase the analyzer's source output power so that the test device output power is
at the top of the measurement range of the analyzer test port.
Press
and enter the new source power level, followed by
Power
CAUTION
Do not exceed the maximum test port power level that is printed on
the front panel of your network analyzer. Exceeding this maximum
power level may damage your analyzer.

Reduce the Receiver Noise Floor

Refer to
"Reducing Noise" on page

Reduce the Receiver Crosstalk

Refer to
"Reducing Receiver Crosstalk" on page
5-16.
5-18.
Optimizing Measurement Results

Increasing Dynamic Range

.
x1
5- 15

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8719et8720es8720et8722es8722et

Table of Contents