Power Supply Section - Agilent Technologies 4263B LCR Service Manual

Table of Contents

Advertisement

The EEPROM (A2U1) stores the unit unique data, such as the adjustment data and the operator
unique data, such as the operator saved measurement setup condition. The EEPROM will
not lose its data when power is turned o . The EEPROM is mounted in a socket for easy
replacement.
Jumpers and switches set the 4263B's conditions. Refer to Appendix B for more information.

Power Supply Section

Figure 4-9 shows the power supply section block diagram.
The ac line power is connected to the A5 AC Inlet Board Assembly, which consists of the fuse,
line switch, and varistors. The voltage selection switch on the rear panel selects the ac line
voltage from 230V or 115V. According to the selection, the ac line power is connected to the
corresponding primary tap of the transformer. Varistor, VR1 protects the circuit from surge
voltage on the line power. VR2 blows fuse F1 if the 4263B is turned on with the wrong line
voltage setting selection, for example, 115V selection when using a 230V line.
The transformer secondary voltage is approximately 14 V. The ac voltage is recti ed and
ltered by circuitry on the A1 Main Board Assembly, and produces approximately 14 Vdc. The
dc voltage is applied to the DC-DC Converter to produce regulated +12V, +5V, and 12 Vdc.
The dc voltages are supplied to all assemblies, either directly or indirectly via the A1 assembly.
4-10