Required Adjustment after Replacing Assembly
After replacing the following assembly, the adjustment items described in Table 3-1 must
be required. The adjustment must be performed Agilent's qualified service personnel. If
you need the adjustment for your E5061A/E5062A, it should be sent to the nearest Agilent
Technologies service office.
Table 3-1
Required Adjustment Item after Replacing Assembly
Source Board
Receiver Board
Hard Disk Drive
Analog Interface Board (PCI DSP Card)
Writing ID
This item writes the serial number and the option structure into the E5061A/E5062A.
Required equipment for the writing ID
None
Frequency Reference Adjustment
The purpose of this procedure is to adjust the 10 MHz frequency reference on the receiver
board.
Chapter 3
Required Adjustment after Replacing Assembly
Replaced Assembly
Adjustment Item
Adjustment
83