To connect the analyzer to an external monitor

The External Monitor connector is a 15-pin connector with standard VGA
pinout. The External Monitor connector can interface with an external,
multi-scanning monitor. The monitor must have a 25.5 kHz horizontal scan rate,
a 60 Hz vertical refresh rate, and must conform to EIA-343-A standards.
l
Connect the analyzer's rear panel EXTERNAL MONITOR connector to an
external monitor using an appropriate cable.
For additional information, see EXTERNAL MONITOR in the analyzer's online
help or the Service Guide.

To connect an external frequency reference

The analyzer may be connected to an external 1, 2, 5, or 10 MHz frequency
reference. The amplitude of the 10 MHz external frequency reference must be
between −3 dB and +20 dB into 50 ohms.
l
Using a BNC cable, connect the external frequency reference to the rear panel
EXT REF IN connector.
The analyzer uses its internal frequency reference if an external frequency
reference or the optional OVEN REF OUT is not connected to EXT REF IN.