Optical Insertion Loss Test - Agilent Technologies 11980A Operation And Service Manual

Fiber optic interferometer
Table of Contents

Advertisement

Description
Equipment
C A U T I O N
1. Note that the Agilent 71400C/71401C lightwave signal analyzer has a built-in optical power

Optical Insertion Loss Test

This test measures the optical insertion loss of the Agilent 11980A at a wave-
length in its 1250 nm to 1600 nm operating range.
Lightwave source. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 83400A
Optical average power meter
Single mode optical fiber patch cords
and adapters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . see
Figure 5-1. Insertion Loss Test Setup
Do not touch the tips of the front-panel input or output connectors. The optical
interface on the Agilent 11980A is composed of precision-polished single-mode
connectors that are easily scratched. Any damage to the optical interface will
increase the insertion loss of the Agilent 11980A and damage the connector
interfaces.
meter.
1
. . . . . . Agilent 8153A (shown) or Agilent 8152A
Operation Verification
Operation Verification
Table 5-1 on page 5-2
5-3