Taking Polarization Dependent Measurements
Measuring Polarization Dependent Loss

Measuring PDL and Insertion Loss Simultaneously

ATTEN TI O N Before you measure Insertion Loss, you must at
least measure and store the reference power P ref. If you have not yet
stored the reference power, refer to page 31 in section 1.6 "Getting
Started".
1 Press A
to activate the "Select Applications" menu.
PPL
2 Choose "PDL / Ins. Loss".
The instrument immediately starts to measure. Wait until the values
are displayed.

Polarization Dependent Loss / Insertion Loss Display

Explanation of the Results

The values are calculated according to the following formulas:
Polarization Dependent Loss
PDL
=
PDL
(dB)
A/B
For most components (except polarizers) the PDL is much smaller
than the Insertion Loss.
44
P
max
10
log
----------- -
P
min
P
max
A/B
=
10
log
----------------------- -
P
min
A/B