Administration Set / Syringe Information (Continued)
The Pump Module uses a wide variety of Pump Module/
FU
RATE (mL/h)
Gemini Infusion System administration sets. The sets
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
are designed for use with the Pump Module as well as for
gravity-flow, stand-alone use.
Primary set must be primed before use. It can be
loaded into Pump Module to deliver a large volume
infusion or it can be set up to deliver a gravity infusion.
• Safety clamp fitment (referred to on device as "Flo-
Stop") is a unique clamping device on the pumping
segment that is part of all Pump Module/Gemini
Infusion System sets (see "Safety Clamp Fitment").
The Syringe Module uses standard, single-use,
disposable syringes (with luer-lock connectors) and
FU
administration sets, designed for use on syringe pumps.
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
• For a list of compatible syringes, see "Compatible
Syringes".
SmartSite
1.
Prior to every access, swab top of Needle-Free Valve port
with
isopropyl alcohol (
70%
(approximately
30
2.
Prime valve port. If applicable, attach syringe to Needle-
Free Valve port and aspirate miniscule air bubbles.
3.
Replace every
72
occurs first. For infusions of blood, blood products or lipid
emulsions, replace every
NOTE:
Dry time is dependent on area temperature, humidity and
ventilation.
Alaris
®
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
Infusion Set (
®
Pump Module
-
seconds) and allow to dry
1
2
seconds).
hours or after
activations, whichever
100
hours.
24
FU
RATE (mL/h)
)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
CAUTIONS
• If the Needle-Free Valve is
accessed by a needle in an
emergency, the valve will be
damaged, causing leakage.
Replace Needle-Free Valve
immediately.
• The Needle-Free Valve is
contraindicated for blunt
cannula systems.
• Do not leave slip luer syringes
unattended.
2-83
General Information