Primary Infusion
IV Fluid Infusion (Continued)
Rate/Volume Infusion (Continued)
•
Syringe Module:
VTBI ALL
♦
When
inactive and estimated available volume in syringe
displays. When
value can be entered.
VTBI ALL
♦
When
pressed, estimated available volume in syringe
displays.
3.
Verify correct infusion parameter entry and press
soft key.
•
If programmed IV Fluid is outside Soft Limit for that
care area, an audio alert sounds and a visual prompt
appears before programming can continue. If Yes soft
key is pressed, programming continues; if No soft key
is pressed, infusion needs to be reprogrammed.
•
If programmed IV Fluid is outside Hard Limit for that
care area, an audio alert sounds and a visual prompt
appears before programming can continue. Infusion
needs to be reprogrammed.
If a Soft Limit is overridden, G icon displays. When G
•
soft key is pressed, all applicable out-of-range limits
are listed.
4.
Syringe Module: Attach administration set to patient.
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
®
Alaris
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
- With Guardrails
Suite MX Protection (Continued)
®
VTBI
is enabled,
soft key is
VTBI
soft key is pressed, a new
VTBI
is disabled and
soft key is
START
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
A
Guardrails Fluid Setup
fat emulsion 20%
PRIMARY INFUSION
0.8
RATE
mL/h
Available:
ALL
VTBI
(19.9 mL)
>Press START
VOLUME
PAUSE
START
DURATION
Midtown Hospital
Adult ICU
G
A
VTBI = 250 mL
VOLUME
AUDIO
INFUSED
ADJUST
Programming
2-41