Cardinal Health Alaris Directions For Use Manual page 105

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

Primary Infusion
IV Fluid Infusion (Continued)
Rate/Volume Infusion (Continued)
Syringe Module:
VTBI ALL
When
inactive and estimated available volume in syringe
displays. When
value can be entered.
VTBI ALL
When
pressed, estimated available volume in syringe
displays.
3.
Verify correct infusion parameter entry and press
soft key.
If programmed IV Fluid is outside Soft Limit for that
care area, an audio alert sounds and a visual prompt
appears before programming can continue. If Yes soft
key is pressed, programming continues; if No soft key
is pressed, infusion needs to be reprogrammed.
If programmed IV Fluid is outside Hard Limit for that
care area, an audio alert sounds and a visual prompt
appears before programming can continue. Infusion
needs to be reprogrammed.
If a Soft Limit is overridden, G icon displays. When G
soft key is pressed, all applicable out-of-range limits
are listed.
4.
Syringe Module: Attach administration set to patient.
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
®
Alaris
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
- With Guardrails
Suite MX Protection (Continued)
®
VTBI
is enabled,
soft key is
VTBI
soft key is pressed, a new
VTBI
is disabled and
soft key is
START
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
A
Guardrails Fluid Setup
fat emulsion 20%
PRIMARY INFUSION
0.8
RATE
mL/h
Available:
ALL
VTBI
(19.9 mL)
>Press START
VOLUME
PAUSE
START
DURATION
Midtown Hospital
Adult ICU
G
A
VTBI = 250 mL
VOLUME
AUDIO
INFUSED
ADJUST
Programming
2-41

Advertisement

Table of Contents
loading