Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

Primary Infusion
Continuous Infusion (Continued)
If Yes was selected and facility has defined a Clinical
Advisory for that drug, a message appears. To
indicate information has been noted and continue
programming, press
If Yes was selected to continue programming, drug
amount and diluent volume (if defined in Drug Library)
are automatically entered for selected drug.
If selected drug had "_ _ / _ _ mL" concentration, drug
amount and diluent volume need to be entered.
If selected drug is not weight-based, Not Used
PATIENT WEIGHT
displays in
If hospital/facility practice guidelines identify selected
drug as weight-based, prompt for a patient weight in
kilograms appears (as in illustrated example, which
reflects use of Alteplase).
4.
Verify parameters are correct and press
confirm.
2-26
Programming
- With Guardrails
Suite MX Protection (Continued)
®
CONFIRM
soft key.
field.
NEXT
soft key to
®
Alaris
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
Alteplase
A
Stroke
Clinical Advisory:
This dosing is for Acute
Ischemic STROKE
>Press CONFIRM
CONFIRM
A
Alteplase
Stroke
100
DRUG
mg
AMOUNT
100
DILUENT
mL
VOLUME
PATIENT
Used
WEIGHT
TIME
hour
UNITS
DOSING
mg/kg/h
UNITS
[Conc]: 1 mg/mL
>Press NEXT to Confirm
DRUG
NEXT
LIBRARY
A
Alteplase
Stroke
100
DRUG
mg
AMOUNT
100
DILUENT
mL
VOLUME
PATIENT
_ _ _ _
kg
WEIGHT
TIME
hour
UNITS
DOSING
mg/kg/h
UNITS
[Conc]: 1 mg/mL
>Enter Patient Weight
DRUG
LIBRARY