Pump Module (Continued)
Alarm
Pump Chamber Blocked
Restart Channel
Syringe Module
Alarm
Occlusion
Pressure Disc Installed
Pressure Disc Removed
2-118
Troubleshooting and Maintenance
Alarms, Errors, Messages (Continued)
Alarms (Continued)
Blocked pump chamber detected.
Door opened and closed during
an infusion. Infusion stops on
affected module.
Module paused for
Increased back pressure sensed
while infusing. Infusion stops on
affected module.
Pressure sensing disc installed
during an infusion. Infusion stops
on affected module.
Pressure sensing disc removed.
Infusion stops on affected module.
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Meaning
minutes.
2
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Meaning
Alaris
Response
Open door and inspect pump
chamber. To open blockage, as
required, massage tubing. To
continue infusion, press
RESTART
key and then
Close door. Press
CHANNEL SELECT
or press
START
and then
soft key.
RESTART
Press
key, or press
CHANNEL SELECT
START
soft key.
Response
Clear occlusion. Press
CHANNEL SELECT
key, or press
START
key and then
CONFIRM
Press
soft key and
RESTART
key.
Reinsert pressure sensing disc and
RESTART
press
key.
®
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
RESET
soft
key.
RESTART
key,
key
key and then
RESTART
soft key.