Primary Infusion
Continuous Infusion (Continued)
7.
Verify parameters are correct and press
If programmed continuous dose infusion is outside
Soft Limit for that care area, an audio alert sounds
and a visual prompt appears before programming can
continue. If Yes soft key is pressed, programming
continues; if No soft key is pressed, infusion needs to
be reprogrammed.
If programmed continuous dose infusion is outside
Hard Limit for that care area, an audio alert sounds
and a visual prompt appears before programming can
continue. Infusion needs to be reprogrammed.
If a dose outside of Soft Limit has been entered and
verified as correct, Message Display also shows either
"
" for a low dose or "
LLL
-- Continued on Next Page --
2-28
Programming
- With Guardrails
Suite MX Protection (Continued)
®
START
↑↑↑
" for a high dose.
soft key.
®
Alaris
System (with v9 Model 8000) DFU
Pump and Syringe Modules Section
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
Alteplase
A A A
Stroke
CONTINUOUS INFUSION
56.7
RATE
mL/h
100
VTBI
mL
0.81
DOSE
mg/kg/h
[Conc]: 1 mg/mL
>Press START
PAUSE
SETUP BOLUS
START
Alteplase
A
Stroke
Dose exceeds
Yes
Guardrail limit of
No
0.81 mg/kg/h.
Proceed?
>Press Yes or No
Alteplase
A
Stroke
Dose exceeds
Reprogram
Guardrails hard
limit of 0.82 mg/kg/h.
>Press REPROGRAM

Advertisement